Assessment of blood lactate measurement accuracy in athletes using three portable analyzers

https://doi.org/10.47529/2223-2524.2026.1.7

Abstract

Objective. To evaluate the accuracy of blood lactate concentration measurements using three portable analyzers compared to a stationary reference instrument.

Materials and methods. Capillary blood samples were collected from highly qualified athletes during staged comprehensive examinations under laboratory conditions. Blood lactate concentrations at rest and following incremental exercise tests on a ski ergometer, cycle ergometer or treadmill were determined using three portable lactate analyzers (Lactate Plus, YST and Eaglenos) and a semi-automatic photometric analyzer (BTS-350).

Results. The limits of agreement for lactate measurements by the Lactate Plus, YST and Eaglenos analyzers compared to the reference instrument exceeded the acceptable interval set at ±2.1 mmol/L. All portable lactometers systematically underestimated lactate concentrations by 2.0–4.0 mmol/L relative to the reference analyzer, which critically distorts the assessment of exercise intensity. Standard errors of estimate of lactate concentration measurements across the entire range were 0.95, 0.63 and 0.91 mmol/L for Lactate Plus, YST and Eaglenos, respectively, and increased with rising blood lactate concentration.

Conclusion. The portable lactometers examined cannot be recommended for accurate blood lactate measurement in athletes under field or laboratory conditions.

About the Authors

M. A. Dikunets
Federal Science Center of Physical Culture and Sport (VNIIFK)
Russian Federation

Marina A. Dikunets, Cand. Sci. (Chem.), Leading Researcher, Biochemistry Laboratory

10, bldg. 1 Elizavetinsky Lane, Moscow, 105005



G. A. Dudko
Federal Science Center of Physical Culture and Sport (VNIIFK)
Russian Federation

Grigory A. Dudko, Senior Researcher, Biochemistry Laboratory

10, bldg. 1 Elizavetinsky Lane, Moscow, 105005



A. A. Arkhipkin
LLC Scientific Center EFIS
Russian Federation

Alexander A. Arkhipkin, Cand. Sci. (Bio.), Head of Clinical Diagnostic Laboratory

16–18, Sadovaya-Chernogryazskaya, Moscow, 1050662



A. S. Kryuchkov
Federal Science Center of Physical Culture and Sport (VNIIFK)
Russian Federation

Andrey S. Kryuchkov, Cand. Sci. (Pedagogy), Head of National Sports Center

10, bldg. 1 Elizavetinsky Lane, Moscow, 105005



A. G. Abalyan
Federal Science Center of Physical Culture and Sport (VNIIFK)
Russian Federation

Avak G. Abalyan, Dr. Sci. (Pedagogy), Associate Professor, Director General

10, bldg. 1 Elizavetinsky Lane, Moscow, 105005



References

1. Wasserman K., Whipp B.J., Koyl S.N., et al. Anaerobic threshold and respiratory gas exchange during exercise. J. Appl. Physiol. 1973;35(2):P. 236243. DOI: 10.1152/jappl.1973.35.2.236.

2. Faude O., Kindermann W., Meyer T. Lactate threshold concepts: how valid are they? Sports Med. 2009;39(6):469-490. DOI: 10.2165/00007256-200939060-00003.

3. Bentley D.J., Newell J., Bishop D. Incremental exercise test design and analysis: implications for performance diagnostics in endurance athletes. Sports Med. 2007;37(7):575-586. DOI: 10.2165/00007256-200737070-00002.

4. Bagger M., Petersen P.H., Pedersen P.K. Biological variation in variables associated with exercise training. Int. J. Sports Med. 2003;24(6):433-440. DOI: 10.1055/s-2003-41180.

5. Raa A., Sunde G.A., Bolann B., et al. Validation of a point-of-care capillary lactate measuring device (Lactate Pro 2). Scand. J. Trauma Resuscit. Emerg. Med. 2020;28(1):83. DOI: 10.1186/s13049-020-00776-z.

6. Bourdon P.C., Woolford S.M., Buckley J.D. Effects of varying the step duration on the determination of lactate thresholds in elite rowers. Int. J. Sports Physiol. Perform. 2018;13(6):687-693. DOI: 10.1123/ijspp.2017-0258.

7. Jamnick N.A., Botella J., Pyne D.B., et al. Manipulating graded exercise test variables affects the validity of the lactate threshold and VO2peak. PLoS One. 2018;13(7):e0199794. DOI: 10.1371/journal.pone.0199794.

8. Ploszczyca K., Jazic D., Piotrowicz Z., et al. Comparison of maximal lactate steady state with anaerobic threshold determined by various methods based on graded exercise test with 3-minute stages in elite cyclists. BMC Sports Sci. Med. Rehabil. 2020;12(1):70. DOI: 10.1186/s13102-020-00219-3.

9. Bonaventura J.M., Sharpe K., Knight E., et al. Reliability and accuracy of six hand-held blood lactate analysers. J. Sports Sci. Med. 2015;14(1):203-214.

10. Mentzoni F., Skaugen M., Eythorsdottir I., et al. Precision and accuracy of four handheld blood lactate analyzers across low to high exercise intensities. Eur. J. Appl. Physiol. 2024;124(12):3781-3788. DOI: 10.1007/s00421-024-05572-6.

11. Tanner R.K., Fuller K.L., Ross M.L.R. Evaluation of three portable blood lactate analysers: Lactate Pro, Lactate Scout and Lactate Plus. Eur. J. Appl. Physiol. 2010;109(3):551-559. DOI: 10.1007/s00421-010-1379-9.

12. Pyne D.B., Boston T., Martin D.T., et al. Evaluation of the Lactate Pro blood lactate analyser. Eur. J. Appl. Physiol. 2000;82(1-2):112-116. DOI: 10.1007/s004210050659.

13. Baldari C., Bonavolonta V., Emerenziani G.P., et al. Accuracy, reliability, linearity of Accutrend and Lactate Pro versus EBIO plus analyzer. Eur. J. Appl. Physiol. 2009;107(1):105-111. DOI: 10.1007/s00421-009-1107-5.

14. Hart S., Drevets K., Alford M., et al. A method-comparison study regarding the validity and reliability of the Lactate Plus analyzer. BMJ Open. 2013;3(2):e001899. DOI: 10.1136/bmjopen-2012-001899.

15. Crotty N.M., Boland M., Mahony N., et al. Reliability and validity of the Lactate Pro 2 analyzer. Meas. Phys. Educ. Exerc. Sci. 2021;25(3):202-211. DOI: 10.1080/1091367X.2020.1865966.

16. Medbo J.I., Mamen A., Holt Olsen O., et al. Examination of four different instruments for measuring blood lactate concentration. Scand. J. Clin. Lab. Invest. 2000;60(5):367-380. DOI: 10.1080/003655100750019279.

17. Rathee K., Dhull V., Dhull R., et al. Biosensors based on electrochemical lactate detection: a comprehensive review. Biochem. Biophys. Rep. 2015;5:35-54. DOI: 10.1016/j.bbrep.2015.11.010.


Review

For citations:


Dikunets M.A., Dudko G.A., Arkhipkin A.A., Kryuchkov A.S., Abalyan A.G. Assessment of blood lactate measurement accuracy in athletes using three portable analyzers. Sports medicine: research and practice. 2026;16(1):12-23. (In Russ.) https://doi.org/10.47529/2223-2524.2026.1.7

Views: 182

JATS XML


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 2223-2524 (Print)
ISSN 2587-9014 (Online)